Simple and rapid docetaxel assay in plasma by protein precipitation and high-performance liquid chromatography-tandem mass spectrometry

J Chromatogr B Analyt Technol Biomed Life Sci. 2004 May 25;804(2):263-7. doi: 10.1016/j.jchromb.2004.01.021.

Abstract

A simple, rapid and low cost sample preparation method was developed for quantification of docetaxel in mouse plasma by high-performance liquid chromatography/tandem mass spectrometry with paclitaxel as the internal standard. A small volume of plasma (40 microl) and one-step protein precipitation using methanol and acetonitrile (1:1 (v/v)) were used for sample preparation. The calibration curve for docetaxel in mouse plasma was linear over the range 25-2500 nM. The detection limit was 8 nM. The lower limit of quantitation is 25 nM. The intra- and inter-day precisions (CV) of analysis were 9.5 and 9.7% for the low quality control (LQC), 5.5 and 4.9% for the medium quality control (MQC) and 3.9 and 6.3% for the high quality control (HQC), respectively. The accuracy was 102.5% for LQC, 97.9% for MQC and 108.8% for HQC. This assay has now been applied to evaluation of mouse pharmacogenetics and other clinical pharmacology applications.